The ARCHITECT i2000SR PLUS immunoassay analyzer meets your laboratory’s high standards by delivering STAT results when you need them. The flexible protocols built into the ARCHITECT i2000SR enhance laboratory workflow and allow you to report results with confidence.
The ARCHITECT i2000SR PLUS immunoassay analyzer offers a maximum throughput of up to 200 tests per hour. Featuring a load-up capacity of 135 samples with 35 priority and 100 routine areas, the ARCHITECT i2000SR has 25 refrigerated reagent positions.
For in vitro diagnostic use only.
Methods | CHEMIFLEX |
Maximum Throughput | Up to 200 tests/hour |
Sample Types | Serum, Plasma, Whole Blood, Urine |
Sample Tubes | Height: 72-102 mm Diameter: 9.6-16.1 mm |
Sample Cup | Yes (50 µL dead volume) |
Sample Capacity | 135 |
Sample Barcode Types | Code 39, Codabar, Interleaved 2 of 5, Code 128 |
Sample Result Storage | 50,000 |
Sample Volume | 10-150 μL, Average: 57 μL |
Automatic Dilution | Yes |
Sample Probe Carryover | ≤0.1 parts per million |
Reagent Capacity | 25 refrigerated positions |
Reagent Type | 100% liquid ready-to-use |
Reagent Onboard Stability | 14-30 days |
Calibration Frequency | Calibrate with new lot number, if controls are out of range or if specified otherwise within the package insert |
Sample, Clot and Bubble Detection | Yes |
Reagent Pressure Monitoring | Yes |
Sample Interference Measurement | No |
System Control Center | 1 SCC, with color touchscreen monitor, keyboard and mouse |
Onboard Maintenance Records | Yes |
Online Error Code Help | Yes |
Host Interface | Bidirectional, serial RS-232 interface, host query option available |
Remote Diagnostics | AbbottLink |
Dimensions (H x W x D) | 48” x 61” x 49” 121.9 x 154.9 x 124.5 cm |
Weight | 1,081 lbs 490.3 kg |
Electrical Requirements | AC 180-264V, 47-63 Hz |
Water Requirements | Purified water to dilute buffer concentrate |
Heat Output* | 4280 BTU/hr, running mode |
Sample Loading | RSH |