Dimension Xpand, SN: 2004081879


Significant dust inside analyser and PC. Replaced CR20232 battery (had 0.4V, date error appeared). Started up without UPS (UPS operational but has faulty batteries). After start-up, communication error and disk error appear.No mechanisms are moving.Monitor touch panel operational. PC probably crashes.


UPS condition (Falcon, SG2K-1T, SN: 0606041023): UPS operational. A check was carried out after connecting working batteries. UPS only starts up with working batteries connected. Batteries are missing. Originally had 6 batteries: EUROPOWER EV 9-12 (12V, 36W/Cell 15 min). UPS originally for Dimension Xpand Plus analyser. NOTE: NO BATTERY CABLING (5 cables terminated with female connectors are missing). Unit vacuumed.


Specifications

ConditionUsed
Stock Number2004081879