Description
Power supply (220-240) VAC (can be changed to (110-120) VAC). 2 new magnets for the reagent mixer and 2 new magnets for the left sample mixer were glued (the magnets of the right mixer remained original). Both the upper and lower parts of the housing are after gluing and painting (MK). The lower, green part of the housing was painted white. Missing legs were supplemented (4 pcs.). The nameplate was recreated. Calibration of the incubator temperature (increase of the block temperature by 0.2 degrees). The analyzer starts correctly, the printer is functional, the display is functional, the keyboard is functional. Service tests and the "Utilities" check were carried out - see below. The analyzer is functional in the scope of the activities performed, in good condition. "Wet" tests were not carried out.
Kit contents:
1. Fibrintimer II (BFT*II)
2. Thermal printer paper roll
3. Paper holder (set, 2+1 elements) - alternative
4. 230VAC power cable.
5. NONE: GW5 reagent bottle adapter for 5 ml reagent bottle, there should be 3 pcs.
6. NONE: measuring socket cap 1 pc. (there is 1 pc. there should be 2 pcs.).
7. NONE: 100 cuvettes (Dispo-System), Teflon reagent mixing paddle, equipment cover, spare fuses, user manual.
Specifications
Manufacturer | SIEMENS |
Model | BFT II |
Condition | Used |
Serial Number | L0160305 |
Stock Number | 9163001215 |