Advia Centaur XPT

Advia Centaur XPT

Location:Przeźmierowo, Poland

orCall +48 601 333 666

Description

Advia Centaur XPT, SN: IRL23121843



Assembly and commissioning of the analyser. It was necessary to connect the internal power supply plug and to connect the temperature block supply cables terminating in meshes to the power supply unit (power supply unit operational). The manifold oviposition was left removed in the central part of the analyser (the oviposition is in the lower part of the analyser).  The analyser was started up with a monitor borrowed from another Centaur XPT analyser. When the analyser is switched on, various messages and errors appear (see photos). No mechanisms move during initialisation. Initialisation ends with the message ‘Unknown’. The kit contained the wrong 19’ monitor with the wrong 4:3 aspect ratio and too low a resolution - this monitor was taken to the main warehouse. A non-touch monitor was added to the set: LOC, E2260SDA, 220LM00010, SN: FDND1HA006221 and a power cable was added (the original monitor is powered by a 12V power supply).  Mechanisms secured with 6 cable ties.


Specifications

ConditionUsed
Stock NumberIRL23121843