Haematology, Coulter T890, SN: N31903
Zasilanie: 120VAC/2,39A/286,8W. Included: complete analyser, cabling and hosing, no 230VAC/115VAC transformer. Dismantling, maintenance of the interior and casing fascia. Replacement of hydraulic valve ovings (approx. 50% ovings). Removal and inspection of fluid sensor. Replacement of rubber fluid sensor grommets with alternative ones - they were causing punctures to the sensor housing. Maintenance of one of the hydraulic valves. Follow-up measurements on patient's whole blood. Results as expected. Flushing out hydraulics and drying analyser. Protection of ceramic valve with spacer and label. Analyser in working order. Hard copy manual available on site (scans available).
A video has been recorded - an attempt will be made to upload this in a subsequent email.