BC-5800
BC-5800
Location:Przeźmierowo, Poland
Description
BC-5800_SN_RW-56003521
Year of manufacture 06.2015. Power supply for analyser 100-240VAC, 300VA. Maintenance, removal of labels. Tests performed: without reagents, without compressor, without reagent connections. Analyser mechanical/electrical/electronic side starts correctly - except for the deficiencies described below. Password: ADMIN/ADMIN. Selected mechanical tests - positive. Temperatures - correct. Power supply voltages - correct. Laser diode current: 32.13 (range 20-50). Language change from Polish to English. Boot software version of analyser: 1.3.5
LACKS:
1. Feeder missing.
2. ‘Pneumatic Unit’ compressor missing.
3. Multiple reagent connections, hoses, effluent sensor with oviposition, power cable missing.
4. CR1220 battery damaged, replacement necessary, analyser loses date and time.
5. 2 screws of the rear sheet metal broken.
6. Pressure connection at rear of analyser missing.
Broken plastic of the front casing. 8.
8. Hand-held barcode reader.
Specifications
Condition | Used |
Stock Number | RW-56003521 |