BC-5800

BC-5800

Location:Przeźmierowo, Poland

orCall +48 601 333 666

Description

BC-5800_SN_RW-53003381



Year of manufacture 03.2015. Power supply for analyser 100-240VAC, 300W. Compressor power supply 220/230VAC. Maintenance, removal of labels. All tests performed without reagents. Analyser mechanical/electrical/electronic side starts correctly. Password: admin/admin.

Test procedure:

1. CR1220 battery voltage check: faulty, replacement necessary, analyser loses date and time.

2. presence of reagents simulated.

3. expiry date of reagents changed.

4. analyser software version - Boot software: 1.3.5

5. Temp&Pres - temperatures and pressures in correct ranges.

6. Vol&Cur - voltages and currents in correct ranges. Laser diode current: 36.3 (range 20-50).

7. Self test: tests (Mechanical, Circuit, Valve) positive except: RBC Aperture Vol, Hgb Blank Voltage, Internal Barcode, Sample Feed, Sample Load, Sample Unload, Pierce Cylinder. Dobber - reagent missing. Errors are probably due to lack of reagents and lack of feeder.

8 Setup: date correction not performed, Print - user data was not present, Advanced (service password) - language change from Polish to English.



Kit composition:

1. analyser

2. compressor, year of manufacture 07.2015, SN: RWA57018626, 220/230VAC.

3. hand-held barcode scanner - NONE.

4. Complete hosing and cabling.

6. sample wands - NONE.

7. feeder (including internal barcode scanner) - NONE.


Specifications

ConditionUsed
Stock NumberRW-53003381