BC-5800, Mindray

BC-5800, Mindray

Location:Przeźmierowo, Poland

orCall +48 601 333 666

Description

BC-5800, Mindray, SN: RW-57003527


Year of manufacture 07.2015. 100-240VAC power supply for analyser. Compressor power supply 220/230VAC.Maintenance, removal of labels.All tests performed without reagents.Analyser mechanical/electrical/electronic side starts correctly.

Test procedure:

1. The presence of reagents was simulated.

2. expiry date of reagents was changed.

3. analyser software version - Boot software:1.3.5

4. Temp&Pres - temperatures and pressures within correct ranges.

Vol&Cur - voltages and currents in correct ranges.Laser diode current: 29.7 (range 20-50).

6. Self test: tests (Mechanical, Circuit, Valve) positive except:RBC Aperture Vol, Hgb Blank Voltage. Dobber - no reagent. Errors are probably due to lack of reagents.

Setup: date correction, Print - delete user data, Advanced - change language from Polish to English.

Kit composition:

1. analyser

2. compressor, year of manufacture 07.2015, SN: RWA57013627, 220/230VAC.

3. hand-held barcode scanner (Newland, NLS-HR100, SN: H55E13566NOM)

4. complete hosing and cabling

6 Sampling wands 6 pcs.

7. feeder, SN: RW-57003527.

Specifications

ManufacturerMINDRAY
ConditionUsed
Stock NumberRW-57003527